New developments based on modern hardware and software design expands the instrument’s usability and measurement performances.
New improved compact mechanics increases significantly the reliability and stability of the instrument. True real time measurement, open hardware architecture for integration of instrument to complex systems, ethernet connection, versatile user interface
Features
- direct and precise measurement of dead time and life time
- surface tension as function of physical time and life time
- available time interval of 5 orders of magnitude (0.001 s to 100 s)
- precise measurement of correct dynamic surface tensions
- direct determination of the hydrostatic pressure via automatic capillary immersion
- easy customized experimental protocols - set modes of measurement for different requrement
- correction of hydrodynamic and viscosity effects of a measurement liquid
- continuously temperature monitoring of the sample
- hardware external bus for instrument integration into lab measurement systems - new modern design with accurate high resolution color capacitive touch display - external large format LCD connected direct to instrument - user interface via touch or keyboard/mouse
- optional software control via Ethernet for system integration
- instrument can be directly integrated into local network environment with internet
access and automatic software update
- surface tension as function of physical time and life time
- available time interval of 5 orders of magnitude (0.001 s to 100 s)
- precise measurement of correct dynamic surface tensions
- direct determination of the hydrostatic pressure via automatic capillary immersion
- easy customized experimental protocols - set modes of measurement for different requrement
- correction of hydrodynamic and viscosity effects of a measurement liquid
- continuously temperature monitoring of the sample
- hardware external bus for instrument integration into lab measurement systems - new modern design with accurate high resolution color capacitive touch display - external large format LCD connected direct to instrument - user interface via touch or keyboard/mouse
- optional software control via Ethernet for system integration
- instrument can be directly integrated into local network environment with internet
access and automatic software update
Instrumental parts
User interface provided touch screen functionalities and optional possibilities to use external mouse/keyboard connected with the instrument via
USB port, at the same time user interface may be use large external LCD, connected direct to instrument. The BPA2S uses modern high quality electronic and
mechanic components. This results in a better measurement
accuracy and measurement stability.
USB port, at the same time user interface may be use large external LCD, connected direct to instrument. The BPA2S uses modern high quality electronic and
mechanic components. This results in a better measurement
accuracy and measurement stability.
The instrument allows the following measurements
- surface tension of liquids
- density estimation
Main features of the software
BPA2S operated as stand-alone instrument. All instrument’s functions are supported without external PC connections. As user interface a touch screen provides functionalities and optional possibilities to use external mouse/keyboard connected with the instrument via USB port. USB-2 interface allows storing the measurement data on a USB. An Ethernet interface allows connection of the instrument directly to a local net. The instrument control from a PC via local net or internet is possible.
Technical Specifications
Technical Specifications
Range of surface and interfacial tension | 10 to 150 mN/m |
Resolution | ± 0.005 mN/m |
Accuracy of surface tension measurement | better than 0.5 % |
Dynamic time range: | Standard capillary 10 ms to 20000 ms Special capillary 0.1 ms to 100000 m s |
Accuracy of density measurement | 2 % |
Display | Graphic color TFT touch display 7’’ resolution 800 x 480 pixel |
Minimum volume test liquid | 1 ml |
Measurement temperature range | 0 – 150°C |
Experimental time: | Standard mode: 20-30 min Fast mode: 4-6 min |
Operation system | Windows Embedded standard WES7 |
Firmware | free update over 1 year after purchase, update via Internet |
Measurement Mode |
- Standard - Constant time - Fast Scan - Increasing flow - Density estimation - Viscosity estimation - Custom mode |
Opereation mode | Stand-alone (touch display, mouse, keyboard) or via VNC-client |
Data storing: |
built-in SSD 64 GB USB Memory stick network storage |
Interface | 1 x Ethernet 3 x USB 2.0 / 1 x USB 3.0 WiFi (option) DisplayPort (DVI/HDMI) res. up to 1920x1200 |
Process control option | Hardware 10 bit data bus |
Enviroment temperature | 10 - 50 °C |
Dimensions Measurement unit / Tripod | 300x400x350 mm |
Weight Measurement unit | 5 kg |
Power requerment: | Measurement unit 24 VDC |
Power supply, external | 100 … 240 AC; to 24 VDC |
Max. power consumption | 15 W |
Extra accessories | capillaries of different diameters |